Logging Personal Noise Dose Meter SL-1256DOS
The personal noise dose meter is test equipment to evaluate the noise condition of working environment by measuring the accumulated noise exposure.
Principle of Measuring Accumulated Noise Exposure
DOSE is a parameter used to quantify noise exposure measured in a period, during turns of % noise exposure relative to 90dBA for 8 hours.
i.e. 100% DOSE = 90dBA for 8 hours
This is known as the Criterion. Other criterion are available such as 100% DOSE = 85dBA for 8 hours to meet legislation in different countries.
If the person stayed in this environment for 4 hours, he would receive 50%DOSE.
Also if the noise level was say 93dBA, 3dB higher, a 100% DOSE would be accumulated in only 4 hours, The above example is for a dose meter with 3dB Exchange Rate, where a 3dB increase in sound level corresponds to a doubling energy. A dose meter with 5dB Exchange Rate is also, available where the sound must increase by 5dB to double the dose as required to meet noise legislation in some countries.
Features
This Noise Dose Meter meets the GB/T 15952 standard.
The Noise Dose Meter measure the frequency weighted noise exposures and peak sound levels simultaneously.
* Simple operation, including direct read–out of results.
* Eight built–in standard dose measurement setups : OSHA80, OSHA90, MSHA-80, MSHA-90, DOD, ACGIH, ISO85 and ISO90.
* Nine user – defined measurement setups.
* A single sound level meter (SLM) setup.
* A single sound level meter (SLM) setup.
* 50 dose (DATA) result stores.
* Download of setups from PC to the meter.
* Keypad lock protects meter against accidental operation.
* Timers support pre–selection of measurement time and duration.
* Use USB data output to connect with PC.
The meter is ideal for quick on–site surveys and for monitoring personal noise exposure in accordance with European ISO or USA OSHA, MSHA, DOD AND ACGIH standards.
The meter is two overlapping measurement ranges : 60 – 130 & 70 – 140dB, A & C RMS weightings, Linear & C Peak weightings, Fast & Slow time weightings, and energy exchange rates (Q) of 3, 4, 5 & 6. In dose measurement mode, it display % dose, % dose projected for an 8 hour period, peak level and measurement duration.
For countries that are subject to European Union regulations or the equivalent ISO standards, the meter measure the daily sound exposure level (LAeq,8hr,Pa2h according to IEC 61252, which is identical with the L EX,8h required by ISO 1999), while for USA OSHA/MSHA regulations they measure the Time Weighted Average level (TWA).
When used as a sound level meter it display sound level, time–averaged (L Aeq) sound level, peak and sound exposure level (SEL).
The meter is automatic run timing facilities controlled by the built–in clock and time–history recording capabilities. Sampling times can be specified between 1s and 1 hour, while storage is 120000 data points.
The meter is able to measure and save up to five specified Ln values (Statistical noise levels). Run duration, start and stop times preset, and other setting via the meter keys or software.
Standard Applied | IEC61252, ANSI S1.25 – 1992 for dose meter and sound exposure meters. IEC 61672-1-2013 Class 2 |
Microphone | 1/2-inch electret condenser microphone with 31-inch integral cable |
Display | Liquid Crystal Display |
Measurement Ranges (Linearity and Indicator ranges at 4KHz) | 60 – 130 : 60 – 130dB (A and C) |
70 – 140 : 70 – 140dB (A and C) | |
Peak Ranges (C–weighted or Linear Peak over the top 40dB of each measurement range) | 60 – 130 : 93 – 133dB Peak |
70 – 140 : 103 – 143dB Peak | |
Frequency Weightings | RMS Detector: A or C |
Peak Detector: C or Z (Linear) | |
Time Weightings | Fast and Slow (RMS detector) |
Exchange Rate | 3, 4, 5 or 6dB |
Stabilization Time | 10 seconds from power on |
Threshold and Criterion Levels (From the setup, predefined in the built – in setup according to applicable standards) | User–defined setups in the following ranges : |
Threshold Level : 70 – 90dB in 1dB steps | |
Criterion Level : 80, 84, 85 or 90dB | |
Criterion Level : 80, 84, 85 or 90dB | At 0.1dB above the top of the selected measurement range |
At 0.1dB above the top of the selected measurement range | Under–Range Indications |
High Level Detector | 115dB |
Keypad Lock | Lock and unlock by pressing key combinations: Up and right arrows to lock; left and down arrows to unlock |
Clock | Real–time clock with calendar |
Memory | Results from 50 measurements can be stored for later viewing and download |
Measurement Control | Measurement Duration : May be set to 5, 10, 15, or 30 minutes or 1, 2, 4, 8, 10, 12, or 24 hours |
Timers | Up to 16 timers(automatic start and stop)can be set(maximum up to one month ahead of measurement time) |
Timer Controlled Started / Stop | Set up from the PC software or meter keypad |
Logging | All measurement parameters values can be logged |
Logging Period (LP) | 1s, 2s, 5s, 10s, 15s, 20s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min or 60min |
Logging Capacity | 120000 values can be stored at any one time |
Setup | Eight built–in (predefined) default setups are included. Nine additional user–defined setups can be stored |
Output | USB interface |
Power Supply | 4×1.5V AAA Size (UM-4)Battery |
Battery Indicator | Symbol indication battery voltage level in 4 steps. Indicator flashes when voltage is insufficient for operation |
Operating Temperature & Humidity | 0 – 50ºC (32 to 122ºF) ,10 – 90%RH |
Storage Temperature & Humidity | -10 – 60ºC(14 to 140ºF) ,10 – 75%RH |
Size | 140x72x34mm |
Weight | Approx. 350g (including batteries) |
Standard Accessories | Main Unit |
External Probe | |
Screwdriver | |
Carrying Case (B04) | |
Operation Manual | |
Optional Accessories | USB Data with Software |
Bluetooth Data Adapter with Software |
Setup | OSHA-80 | OSHA-90 | MSHA-80 | MSHA-90 | DOD | ACGIH | ISO-85 | ISO-90 | User1-9 | SLM |
Measurement Range (dB) | 70–140 | 70–140 | 70–140 | 70–140 | 70–140 | 70–140 | 70–140 | 70–140 | 70–140 | 70–140 |
Time Weighting | Slow | Slow | Slow | Slow | Slow | Slow | Fast | Fast | Fast | Slow |
Frequency Weighting | A | A | A | A | A | A | A | A | A | A |
Peak Frequency Weighting | Lin | Lin | Lin | Lin | Lin | Lin | C | C | C | Lin |
Exchange Rate | 5 | 5 | 5 | 5 | 4 | 3 | 3 | 3 | 3 | 3 |
Threshold (dB) | 80 | 90 | 80 | 90 | 80 | 80 | 70 | 70 | 80 | N/A |
Criterion Level (dB) | 90 | 90 | 90 | 90 | 85 | 85 | 85 | 90 | 90 | N/A |
Allow User to Change Setup | No | No | No | No | No | No | No | No | Yes | Yes |
OSHA-80 | OSHA-90 | MSHA-80 | MSHA-90 | DOD | ACGIH | ISO–85 | ISO–90 | User1-9 | SLM |
DOSE% | DOSE% | DOSE% | DOSE% | DOSE% | DOSE% | DOSE% | DOSE% | DOSE% | SPL |
PDOSE% | PDOSE% | PDOSE% | PDOSE% | PDOSE% | PDOSE% | PDOSE% | PDOSE% | PDOSE% | MAX |
TWA | TWA | TWA | TWA | TWA | LEPd | LEPd | LEPd | TWA | LEQ |
PTWA | PTWA | PTWA | PTWA | PTWA | LEQ | LEQ | LEQ | PTWA | LEQ |
LAVG | LAVG | LAVG | LAVG | LAVG | SEL | SEL | SEL | LAVG | PKZ |
LEQ | LEQ | LEQ | LEQ | LEQ | SE(Pa2h) | SE(Pa2h) | SE(Pa2h) | LEPd | |
SEL | SEL | SEL | SEL | SEL | MAX | MAX | MAX | LEQ | |
MAX | MAX | MAX | MAX | MAX | PKZ | PKZ | PKZ | SEL | |
PKZ | PKZ | PKZ | PKZ | PKZ | L10 | L10 | L10 | SE(Pa2h) | |
L10 | L10 | L10 | L10 | L10 | L50 | L50 | L50 | MAX | |
L50 | L50 | L50 | L50 | L50 | L90 | L90 | L90 | PKZ | |
L90 | L90 | L90 | L90 | L90 | L95 | L95 | L95 | L10 | |
L95 | L95 | L95 | L95 | L95 | L99 | L99 | L99 | L50 | |
L99 | L99 | L99 | L99 | L99 | SPL | SPL | SPL | L90 | |
SPL | SPL | SPL | SPL | SPL | L95 | ||||
99 | |||||||||
SPL |