MIC-2000D multipurpose surgical operating microscope
is a double binocular microscope. The main microscope has zoom magnification system、foot control switch, it can bring the clear image, the instrument is available and flexible for ophthalmology operation.
Features
◆ Apochromatic optical balance technology is used in the optical design to provide good binocular fusion, strong stereoscopic sense and good color restoration.
◆ Space Balancing System in used in the instrument to make the microscope lens and arm to push and pull in any direction with ease.
◆ Module design is used in electrical design to provide the instrument with fault self-diagnosis function and simple maintenance.
◆ Good extensibility to add teaching microscope, image device and image system.
Objective focus range (mm) | Standard F=215 | ||||||
Working distance (mm) | 170mm | ||||||
Total MAG. | Main lens | 4.6x~27x | Assistant lens | 6x / 10x / 14x | |||
N.A. | 0.014~0.043 | 0.021 / 0.033 / 0.0375 | |||||
Filed of view diameter(mm) | 4.6x—F46, 27x—F8.5 | ||||||
Main lens zoom ratio | 1:6 | ||||||
Coaxial illumination | Range | ≥F45(F=215mm) | |||||
Model | 6° Cold light Coaxial illumination | 26° Oblique illumination | |||||
Fine focus range | ≥40mm | ||||||
Speed of fine focus (adjustable) | ≤2mm/s | ||||||
X-Y moving range)
|
50mm×50mm
|
X- ← 0~-25;Home position0;X+ → 0~+25 | |||||
Y- ↓ 0~-25;Home position0;Y+ ↑ 0~+25 | |||||||
X-Y moving speed(adjustable) | ≤2mm/s | ||||||
Filter
|
Inner fixed | UV、IR:
wavelength=700~850nm,transmittance ratio T≤1% wavelength=415~680nm,transmittance ratio T≥90% |
Move infrared, ultraviolet light, alleviating damage from the light to the patients | ||||
Optional | Black point filter: no transmitting in the middle Φ2.5 | Form black point in the light path, avoiding burning to the retina | |||||
No red filter wavelength=415~580nm,transmittance ratio T≥90%
wavelength=595~680nm,transmittance ratio T≤1% |
Move all red light over 580nm wavelength, decreasing the influence to the image clearance caused by the blood red color | ||||||
GG435:
wavelength=350~435nm,transmittance ratio T≤1% wavelength=425~675nm,reflectance ratio R≤0.5% |
Decreasing the short wavelength light energy, avoiding light induced corneal injury | ||||||
GG475:
wavelength=350~475nm,transmittance ratio T≤1% wavelength=425~675nm,reflectance ratio R≤0.5% |
Decreasing the blue light energy, avoiding light induced retinitis. | ||||||
CCD optional adapter | 1/3CCD, standard C interface |
Input Voltage | AC220V ±22V/50Hz±1Hz or AC110V±11V/ 60Hz±1Hz |
Power Consumption | ≤170VA |
Temperature | 5℃~40℃ |
Humidity | 30%~75% |
Pressure | 700hPa~1060hPa。 |
评价
目前还没有评价